Portable Hg concentration in the air measuring instrument Model 431   ´ë±âÁß ¼öÀº ³óµµ ÃøÁ¤±â

ultraprecision (0.013ug/m3) measuring instrument Model J 405

Application
  • Mercury Surveys and Soil Screening
  • Spill Response
  • Ensure Worker Safety
  • Research Projects for Stack, Flue and Natural Gas
  • Hazardous Waste Sites
  • Mercury Exclusion Tests
  • Monitor Disposal and Recycling of Fluorescent Lamps
  • Exhaust Duct Analysis

Specificatio
Resolution 0.001 mg/m3    (405 0.013ug/m3)
Detection Range 0.003-0.999 mg/m3 (405 999 ug/m3)
Precision 5% Relative Standard Deviation at 0.100 mg/m3
Accuracy ± 5% at 0.100 mg/m3
Response Time 13 s in Sample Mode; 4 s in Survey Mode
Flow Rate 750 cc/min
Power Requirements 100-120 V ~ 50/60 Hz, 1 A or 220-240 V ~ 50/60 Hz, 1 A
Internal Battery Pack Rechargeable nickel-cadmium
Environmental Range 0-40 °C, noncondensing, nonexplosive
Interfaces RS-232 PC using JCI software
Dimensions 6¡± W x 13¡± L x 4¡± H
Weight 7 lb
Warranty One year, factory parts and labor
Certifications European Communities (CE) for 220-240 V~ model only

         

 Àç·á³» ¼öºÐ ÇÔ·® ºÐ¼® ÀåÄ¡ Model Computrac 1000XL 2000XL 4000 XL

2009ºÎÅÍ »õ·ÎÈ÷ Ãâ½ÃµÈ 4000 XL¼öºÐÃøÁ¤ ±â´Â 1000°ú 2000ÀÇ ±â´ÉÀ» º¸°­ÇÑ Á¦Ç°ÀÔ´Ï´Ù- ¾Æ·¡ ºñ±³Ç¥ ÂüÁ¶

Water content measurement, solids content measurement , ash content measurement Model Max 5000

MAX 5000 Specificatio
Sample Size 200 mg to 100 g
Resolution Moisture: 0.0001%; Balance: 0.0001g
Moisture/Solids Range 0.1% to 99.9%
Dry Weight Range 0-300%
Results Display % Moisture, % Solids, % Ash or % Dry Weight
Ending Criteria Predict (Auto or Manual), Rate, Time, Rate of Time or Reliability
Heating Range 25 - 600°C, Set in 1°C increments and maintained to ±1°C within the range of 50 - 300°C; ±2°C within the range of 300 - 600°C
Test Parameter Memory 250 Programs
Program Identification Alphanumeric; programmable prompts for lot or product ID numbers
Test Result Memory Data from last 1,000 tests and graphs from last 100 tests
Statistical Calculation Mean, Standard Deviation, and Relative Standard Deviation
Power Requirements 100-120V ~ 50/60 Hz, 1 amp (standby) or 8 amps (heat on)
Interfaces RS-232 (9-pin standard); parallel printer (LPT - PC standard);
RJ-45 (Ethernet)
Dimensions 470mm L x 323mm W x 24mm H; 18.5" L x 12".5 W x 10" H
Weight 13.6 kg; 30 lbs
Warranty Two years, factory parts and labor (one year international)
Certifications cULus; CE
Accessories Multi-Point Temperature Calibration Module; Printer; Sample Basket; Validation Package; Step Down Power Transformer

 2010 New Version  -Jerome® J605  Hydrogen Sulfide Analyzer - ppb & ppt level H2S Gas Analyser

ÈÞ´ëÇü  H2S

ÃÖÀú³óµµ 20 ppt (0.02 ppb)Á¤¹Ðµµ·Î ºÐ¼®

´ë±âÁß ¾ÇÃë ºÐ¼® ³ó¾÷ºÐ¾ç ³ó¾÷ ±â»ó Åä¾ç¿À¿° ºÐ¾ß

 

Applications:
Features:
Additional Features:
Jerome® J605 Specifications:
Resolution 20 ppt in Range 0
Detection Range 3 ppb -10 ppm
Accuracy/Precision Range 0: ±1 ppb at 5 ppb/10%RSD
Range 0: ±3 ppb at 50 ppb/5%RSD
Range 1: ±0.03 ppm at 0.5 ppm /5%RSD
Range 2: ±0.3 ppm at 5.0 ppm/5%RSD
Response Time
Manual Range Selection  
Range 2 (1.0 to 10.0 ppm) 12 seconds
Range 1 (0.10 to 1.0 ppm) 18 seconds
Range 0 (3 to 100 ppb) 12 to 27 seconds
Auto Range Mode 12 to 52 seconds
Flow Rate 150 ml/min or 0.15 l/min
Power Requirements 12V DC/power adapter
Internal Battery Pack Rechargeable nickel metal hydride (NiMH)
Environmental Range 0-40°C, noncondensing, nonexplosive
Interface USB
Dimensions 11" W x 6" L x 6.5" H/ 28 cm L x 16 cm W x 17 cm H
Weight 5.4 lbs / 2.5 kg
Warranty 1 year, factory parts and labor
Certifications European Communities (CE) and TUV: UL 61010-1:2004 standard and CSA-C22.2 NO. 61010-1-04
Options: Functional Test Module for instrument verification in the field